The BP series testing system provides a basic, convenient, fast, and stable platform for performance evaluation and characterization testing of core materials and key components in PEM (Proton Exchange Membrane) electrolysis water hydrogen production. The series BP can achieve performance testing, sensitivity testing, lifespan assessment, and theoretical research of PEM electrolysis cells. Through a professionally developed operating software, the system can control, monitor, and display various parameters and operating conditions of the electrolysis cell in real-time, including water temperature, pressure, flow rate, voltage, current, and other parameters.

Achievable testing functions

  • Testable single cell active area: 1~50 cm²
  • Monitoring of cell current and voltage
  • Polarization curve testing capability
  • Durability testing
  • Sensitivity testing
  • Water circuit temperature, flow, and pressure control
  • Automatic water replenishment function
  • Water circuit conductivity monitoring
  • One-click startup and automatic operation
  • Custom script functionality
  • Employing a two-stage separation technology combining vertical gravity and condensation to achieve efficient gas-liquid separation
  • Equipped with three-stage constant temperature control technology, ensuring the temperature accuracy entering the electrolyzer is maintained within ≤±1°C in steady-state conditions

Product Advantages and Features

  • Compact design, easy to operate.
  • Ready for testing immediately upon startup, with no long waiting time required.
  • Offers the most stable and reliable performance, meeting customers’ long-term testing needs.
  • Equipped with multi-stage automatic safety protection features.
  • The power supply features multiple operating modes, including CC, CV, and CP, as well as protection functions such as OCP, CVP, and OPP.
  • Exclusive upper-level computer operation software with powerful monitoring capabilities.
  • Customizable scripts for simple and convenient operation.
  • Capable of data recording and analysis, facilitating post-test data processing.

ContentParametersNotes
Single-channel Power Supply Power≤500W 
Single-channel Current Range0~100A (Resolution: 10mA) 
Single-channel Voltage Range0~20V(Resolution:1mV) 
Current Setting Precision≤0.1%+0.1%F.S 
Voltage Setting Precision≤0.05%+0.05%F.S 
Single-channel Anode Recirculating Water Flow Rate40~400ml/min 
Recirculating Water Flow Control Precision2.5 
Anode Recirculating Water TemperatureRT+5~90℃  
Conductivity Control≤0.3um/cm 
Temperature Control Precision≤±1℃ 
Preheating Rate≥5℃/min 
Single-channel Maximum Hydrogen Flow Rate Range0.7L/min 
Single-channel Maximum Oxygen Flow Rate Range0.35L/min 
Cathode Backpressure RangeAmbient Pressure 
Ambient Operating Temperature0~45℃ 
Communication InterfaceLAN 
Communication ProtocolModbusTCP、ModbusRTU、SCPI、S7 
Number of Channels1CH4CH8CH32CH 
Model NumberBP-1CBP-4CBP-8CBP-32C 
Power Supply220V/50HZ380V/50HZ380V/50HZ380V/50HZ 
“Equipment Dimensions
(Width × Depth × Height, mm)”
~360*520*4501000*1300*19001500*1300*19004000*2020*2050Subject to adjustment based on actual needs.